PR 9100 Series Manual page 12

Temperature/ma converter
Table of Contents

Advertisement

Fuse ............................................................................... 400 mA SB / 250 VAC
Isolation - test / working:
Input to any ...................................................... 2.6 kVAC / 300 VAC reinforced
Analogue output to supply ........................ 2.6 kVAC / 300 VAC reinforced
Status relay to supply .................................. 1.5 kVAC / 150 VAC reinforced
Communications interface ................................... Communication enabler 4511 /
Signal / noise ratio .................................................. Min. 60 dB (0...100 kHz)
Average response time incl. delay:
Temperature input ......................................... ≤ 1 s
mA input ............................................................ ≤ 0.4 s
Accuracy, the greater of the general and basic values:
Input
type
All
Input
type
mA
Pt100, Pt200,
Pt 1000
Pt500, Ni100,
Ni120, Ni 1000
Pt50, Pt400, Ni50
Pt250, Pt300
Pt20
Pt10
TC type:
E, J, K, L, N, T, U
TC type: R, S, W3,
W5, LR
TC type: B
160...400°C
TC type: B
400...1820°C
10
General values
Absolute
accuracy
≤ ±0.1% of span
basic values
Basic
accuracy
≤ ±16 µA
≤ ±0.2°C
≤ ±0.3°C
≤ ±0.4°C
≤ ±0.6°C
≤ ±0.8°C
≤ ±1.4°C
≤ ±1°C
≤ ±2°C
≤ ±4.5°C
≤ ±2°C
9113 - Product Version 9113-004
Programming front 4501
Temperature
coefficient
≤ ±0.01% of span / °C
Temperature
coefficient
≤ ±1.6 µA / °C
≤ ±0.02°C / °C
≤ ±0.03°C / °C
≤ ±0.04°C / °C
≤ ±0.06°C / °C
≤ ±0.08°C / °C
≤ ±0.14°C / °C
≤ ±0.1°C / °C
≤ ±0.2°C / °C
≤ ±0.45°C / °C
≤ ±0.2°C / °C