7.11 Parts counting

Before the balance can count parts, it must know the average part weight (i.e.
reference). Proceed by putting on a certain number of the parts to be counted. The
balance determines the total weight and divides it by the number of parts, the so-
called reference quantity. Counting is then carried out on the basis of the calculated
average piece weight.
As a rule:
The higher the reference quantity the higher the counting exactness.
In weighing mode
, press and hold until the
message „P 10" appears that is used to set the reference
quantity.
Use
to set the desired reference quantity (such as
100), options include P 10, P 20, P 50, P100, P 200.
Place as many items to be counted (such as 100 items)
as demanded by the set reference quantity and confirm
by
. The weighing scales calculate the reference
weight. The current quantity (such as 100 items) will be
displayed.
Remove reference weight. The balance is from now in
parts counting mode counting all units on the weighing
plate.
Back to Weighing mode by
.
EOC-BA-e-1920