Safety
(Continued)
3.2
Safety instructions
As a matter of principle, the Labomed AXIUS Column is not an electric medical device. According
to the intended use, the AXIUS is a „bearing system for devices, both medical and non -medical,
which are used for dental treatment at the working space of a dentist".
Devices that require connection to an electric supply (low voltage, mains voltage) can also be
attached to the column. Compliance with the following instructions is required for electrical
safety reasons if such devices are attached to the AXIUS Column:
The electric cables must be routed appropriately in the AXIUS to ensure the necessary
freedom of movement required for the motions of the attached device while safeguarding
the cables from damage.
 On principle, there must not be any electrical insulating points inside the AXIUS Column.
The cables routed in the AXIUS Column must be equipped with double insulation in the
column area.
 Cables, whose insulation is damaged, must be replaced!
At the bottom of the AXIUS Column, the attached electric device must be connected with the
equipotential bonding.
 At the bottom of the AXIUS Column, the equipotential bonding must be connected to the
equipotential bonding of the voltage source.
 The devices connected to the AXIUS Column must be included in the safety checks of the
working space of the dentist.
 The specifications of IEC 62353 apply to the electrical safety testing of medical electric
devices or medical electric systems prior to commissioning, during servicing, inspection,
maintenance and following repair or in the case of performance of the statutory prescribed
recurrent tests.
Axius
Axius
9
Part No: 6137120-795
Issue: 1.0
Printed On: June 10, 2023