The ACOM 06AT tuning memory dump file generation and download procedure is performed as follows:
a)
Turn OFF the amplifier. Connect a serial interface cable between the amplifier rear-panel
RS-232 connector and the PC. See the amplifier manual for details on serial cable connection;
b)
With the amplifier powered OFF, launch the ACOM Terminal S program from the PC and:
- Press the CONNECTING SETTINGS button and select a COM port available on your PC.
Configure the connection settings at 9600 bps / 8 bit / No parity / 1 stop bit.
Double check the connection settings before the next step!;
- Press the CONNECT button to open the serial port;
- Do NOT press the FIRMWARE UPDATE button during this procedure!;
c)
Turn on the amplifier to STBY mode and assign the tuner (if not assigned already - see Section
4.2 Tuner Assignment and Unassignment
d)
Call up the "ATU Service" menu on the amplifier's display (see Section
- ATU SERVICE Menu
e)
Go to the "MEMORY DUMP IN A FILE (wait 5-10 min)" on the amplifier screen and press the
SELECT
f)
A new DOWNLOAD window will open in ACOM Terminal S. In it click on the BROWSE button
and select location and name for saving the newly generated dump file on your computer;
g)
Click the SAVE button. The terminal program will then return to the previous window, in which
the DOWNLOAD button should be clicked to start the download of the dump file;
h)
A progress bar will appear, showing the state of the download. It can take 5-10 minutes before
the progress bar is full and the following message is displayed:
Page 42 of 70 | S e c t i o n ADVANCED FUNCTIONS AND OPTIONS

Figure 6-1 | ACOM Terminal S screenshot

Figure 6-1 | ACOM Terminal S screenshot
);
button to confirm the selection;
Operation completed successfully;