46
CLEANING AND DISINFECTING (CONT.)
DISINFECTING
i-STAT ALINITY INSTRUMENT
Disinfection is recommended between each patient. When instrument is dedicated
to a single patient, disinfect at least once a day. The disinfection process must begin
IMMEDIATELY after the cleaning procedure is complete. Standard precautions
should be taken whenever working with blood or blood products.
1
Remove a wipe from the
container and squeeze out
any excess fluid before you
begin to wipe the surface.
3
Allow all
surfaces to
remain wet
for 3 minutes.
SENSITIVE AREAS
Avoid forcing liquid into these areas:
i-STAT Alinity Instrument
Cartridge Port
A
10-Pin Connector under the camera
B
C
Gold contacts (2) on the outside of
the battery
SECTION 8
SECTION 7
SECTION 8
Wipe all surfaces three times.
This includes front, back, sides,
top and bottom.
i-STAT Alinity — Quick Reference Guide
2
3
X
3X
W IPE
W
I
P
E
3X
W IPE
3
X
4
Wipe with gauze
until dry.
B
C
Art: 731848-01
Rev. M
W IPE
3X
3X
W I PE
A
Rev. Date: 25-Jul-2022