44
HOW TO PERFORM PATIENT TESTING (CONT.)
BEST PRACTICE EXAMPLE
i-STAT CHEM8+
Pt: 123456
Na, mmol/L
129
138 - 146 Ref
K, mmol/L
3.0
3.5 - 4.9 Ref
Cl, mmol/L
79
98 - 109 Ref
TCO2, mmol/L
19
24 - 29 Ref
Silence
Use AlinIQ CWi to set reference and action ranges.
RESULTS
® An audible cue will be heard when results are ready.
Touch Silence or remove cartridge to silence the audio.
® When the instrument is customized by the system administrator, reference
ranges as well as action ranges may be assigned.
® When reference ranges are assigned, they appear under the analyte result
followed by "Ref".
® When action (critical) ranges are assigned, results within this range will be
highlighted by both color and an arrow.
¯ Yellow in the result area indicates that the result is outside of the reference
range, but is not within the action range, sometimes known as an abnormal
result. The arrow indicates if the result is high (
¯ Red in the result area indicates that the result is within the action (critical)
range. The arrows indicate if the result is high (
¯ Red arrow in page button - indicates one or more results on second page
are within the action (critical) range.
¯ Yellow arrow in page button - indicates one or more results on second
page are outside the reference range, but not in the action (critical) range.
¯ White arrow in page button - indicates all results on second page are
within the reference range.
SECTION 7
SECTION 6
16FEB2025 09:45
View Second Page
BUN, mg/dL
119
8 - 26 Ref
Crea, mg/dL
6.2
3.5 — 4.9 Ref
i-STAT Alinity — Quick Reference Guide
Op�ons
Menu
Glu, mg/dL
468
75 - 105 Ref
iCa, mmol/L
1.2
1.12 — 1.32 Ref
AnGap, mmol/L
32
-10 — 99 Ref
Page
) or low (
) or low (
Art: 731848-01
Rev. M
Op�ons
Menu
View
Entered Info
Print
Transmit
).
).
Rev. Date: 25-Jul-2022