Download Print this page

Abbott i-STAT Alinity Quick Reference Manual page 38

Hide thumbs Also See for i-STAT Alinity:

Advertisement

34
HOW TO PERFORM QUALITY TESTING (CONT.)
LIQUID QUALITY CONTROL (CONT.)
BEST PRACTICE EXAMPLE
Use AlinIQ CWi to customize for automatic pass/fail
determination using eVAS.
Once customized, the results will appear as follows:
Numeric value with no arrow – In Range
Numeric value with
Numeric value with
i-STAT CHEM8+
i-STAT TriControl L2: 321068
PASS
Na, mmol/L
129
127 - 136
K, mmol/L
3.8
3.5 - 4.1
Cl, mmol/L
92
85 - 98
TCO2, mmol/L
18
16 - 29
Home
i-STAT CHEM8+
i-STAT TriControl L1: 321054
FAIL
Na, mmol/L
124
117 - 126
K, mmol/L
3.0
2.6 - 3.2
Cl, mmol/L
72
68 - 77
TCO2, mmol/L
10
8 - 26
Home
SECTION 6
arrow – Out of Range High
arrow – Out of Range Low
16FEB2025 11:01
BUN, mg/dL
10
8 - 14
Crea, mg/dL
1.3
0.7 - 1.5
16FEB2025 09:45
BUN, mg/dL
66
47 - 62
Crea, mg/dL
3.7
3.0 - 4.6
i-STAT Alinity — Quick Reference Guide
Op�ons
Menu
Glu, mg/dL
Op�ons
91
Menu
85 - 100
iCa, mmol/L
View
1.33
Entered Info
1.24 - 1.42
AnGap, mmol/L
Print
34
Transmit
Page
Op�ons
Menu
Glu, mg/dL
Op�ons
280
Menu
228- 312
iCa, mmol/L
View
0.80
Entered Info
0.73 - 0.89
AnGap, mmol/L
Print
42
Transmit
Page
Art: 731848-01
Rev. M
Rev. Date: 25-Jul-2022