HOW TO PERFORM QUALITY TESTING (CONT.)
LIQUID QUALITY CONTROL (CONT.)
i-STAT CHEM8+
i-STAT TriControl L1: 321055
Na, mmo/L
129
K, mmol/L
3.0
Cl, mmo/L
79
TCO2, mmo/L
19
Home
RESULTS
Use Value Assignment Sheet to determine if results are in range. Follow hospital
policy if results are outside the assigned range.
® An audible cue will be heard when results are ready. Touch Silence or remove
cartridge to silence the audio.
® The result page shown here is the default. The
system administrator must specify the ranges used
in your facility.
® The blinking page button at the bottom of the screen appears when there is more
than one page of results. All action tabs are inactive until the second page of
results has been viewed.
® Occasionally numeric results will be replaced with
the following symbols. When displayed, a new test
must be performed.
<> – Instrument cannot calculate the result.
*** – (Starouts) Instrument is unable to determine a result.
A sample may also yield results that are preceded by a greater than (>) or less
than (<) symbol. These results are outside the instrument's measurement range.
i-STAT Alinity — Quick Reference Guide
SECTION 6
View Second Page
BUN, mg/dL
Glu, mg/dL
119
468
Crea, mg/dL
iCa, mmo/L
6.2
1.2
AnGap, mmo/L
32
Art: 731848-01
Rev. M
Rev. Date: 25-Jul-2022
Op�ons
Menu
Op�ons
Menu
View
Entered Info
Print
Page
33