TROUBLESHOOTING AND SUPPORT
TROUBLESHOOTING
The i-STAT Alinity is programmed to perform quality checks throughout the
testing cycle.
The instrument has several methods of notifying operators of failed quality
checks.
1. Quality Check Failures
- Are displayed when the instrument identifies a problem while running a
cartridge or simulator
- There are 4 types of quality check failures:
1. Instrument
2. Cartridge
3. Sample
4. Software
- Screen displays the type of failure and instructions for resolution.
i-STAT CHEM8+
Pt: 654321
Cartridge Quality Check Failure
Cartridge Was Overfilled
Home
i-STAT CHEM8+
Pt: 654321
Cartridge Quality Check Failure
Cartridge Was Overfilled
When filling a cartridge, use care to
advance blood to the level indicated by
the 'fill to' arrow.
Repeat testing with a freshly filled
cartridge.
Carefully observe the help provided
throughout the testing pathway.
If the same quality check failure displays,
contact the system administrator for
further instruction.
Home
i-STAT Alinity — Quick Reference Guide
SECTION 9
13OCT2025 08:33
Excess blood added to cartridge.
When filling this cartridge, the blood
advanced past the level indicated by
the 'fill to' arrow.
View Resolution
13OCT2025 08:33
Resolution
View Cause
Art: 731848-01
Rev. M
Options
Menu
Code: 37-01
Cause
Options
Menu
Code: 37-01
Rev. Date: 25-Jul-2022