Download Print this page

Abbott i-STAT Alinity Quick Reference Manual page 45

Hide thumbs Also See for i-STAT Alinity:

Advertisement

HOW TO PERFORM PATIENT TESTING (CONT.)
i-STAT CHEM8+
Pt: 123456
Na, mmol/L
129
K, mmol/L
3.1
Cl, mmol/L
81
TCO2, mmol/L
9
When the test is complete, the test results are displayed as in the example
above.
RESULTS
® An audible cue will be heard when results are ready. Touch Silence or remove
cartridge to silence the audio.
® The result page shown here is the default. The system administrator must
specify the ranges used in your facility.
® The blinking page button at the bottom of the screen appears when here is more
than one page of results. All action tabs are inactive until he second page of
results has been viewed.
® Occasionally, numeric results will be replaced with the following symbols.
When displayed, a new test must be performed.
<> – Instrument cannot calculate the result.
*** – (Starouts) Instrument is unable to determine a result.
A sample may also yield results that are preceded by a greater than (>) or less
than (<) symbol. These results are outside the instrument's measurement range.
In order to determine the exact numeric result, the sample must be analyzed by
a different method.
i-STAT Alinity — Quick Reference Guide
SECTION 7
SECTION 6
16FEB2025 09:45
View Second Page
BUN, mg/dL
48
Crea, mg/dL
3.8
Art: 731848-01
Rev. M
Op�ons
Menu
Glu, mg/dL
274
iCa, mmol/L
0.94
AnGap, mmol/L
42
Page
Rev. Date: 25-Jul-2022
41
Op�ons
Menu
View
Entered Info
Print

Advertisement