HOW TO PERFORM QUALITY TESTING (CONT.)
ELECTRONIC SIMULATOR (CONT.)
PRECAUTIONS AND LIMITATIONS
For best results, observe the following precautions:
DO:
• use only the accessories and consumables specified or supplied for this system
by Abbott Point of Care.
• place the removable cap over the contacts when the Electronic Simulator is not
in use or when cleaning or disinfecting.
• check with authorities for local, state, and/or national requirements for disposal.
DO NOT:
• attempt to remove the Electronic Simulator while a test is being performed.
• drop the Electronic Simulator, as doing so can cause it to malfunction.
• place the Electronic Simulator in an oxygen-enriched atmosphere.
• make any unauthorized repairs or modifications to this product.
• use the Electronic Simulator with any instrument other than the i-STAT Alinity.
• touch the area under the cap.
NOTE:
• Function provided by this equipment may be impaired if used in a manner not
specified by Abbott Point of Care.
• The instrument can become contaminated with blood during use. Operators
should use standard precautions whenever handling the instrument,
cartridge, and peripherals to protect themselves from blood-borne pathogens
and pathogens from other body substances. Standard precautions, such as
the wearing of gloves, are designed to protect personnel from blood-borne
pathogens and pathogens from other body substances. These precautions
are based on the assumption that blood, body fluids, and tissue can contain
infectious agents and should therefore be treated as biohazardous materials.
• The instrument and its peripherals are not listed by any authority with respect
to suitability for use in oxygen-enriched atmospheres.
i-STAT Alinity — Quick Reference Guide
SECTION 6
Art: 731848-01
Rev. M
Rev. Date: 25-Jul-2022
27