6) SPECIFICATIONS

General Specifications
Display : 3-3/4 digits 4000 counts LCD display(s)
Update Rate : 3 per second nominal
Polarity : Automatic
Low Battery : Below approx. 2.4V
Operating Temperature : 0
Relative Humidity : Maximum relative humidity 80% for temperature up to 31
decreasing linearly to 50% relative humidity at 40
Altitude : Operating below 2000m
Storage Temperature : -20
Temperature Coefficient : nominal 0.15 x (specified accuracy)/
o
-40
C), or otherwise specified
Sensing : Average sensing for Model 161; True RMS for Model 162
Safety : Meets IEC61010-2-032(2002), EN61010-2-032(2002), UL61010B-2-032(2003)
Category III 600 Volts ac & dc.
Transient protection : 6.5kV (1.2/50µs surge) for all models
Pollution degree : 2
E.M.C. : Meets EN61326(1997, 1998/A1), EN61000-4-2(1995), and
EN61000-4-3(1996)
In an RF field of 3V/m:
Capacitance function is not specified
Other function ranges: Total Accuracy = Specified Accuracy + 45 digits
Performance above 3V/m is not specified
Overload Protections :
Clamp-on jaws : DC 1000A or AC 800A rms continuous
+ & COM terminals (all functions) : 600VDC/VAC rms
Power Supply : standard 1.5V AAA Size (NEDA 24G or IEC R03) battery X 2; or 1.5V
AAA Size (NEDA 24A or IEC LR03) alkaline battery X 2
Power Consumption : typical 11mA for ACA/DCA and 2.9mA for other functions
APO Timing : Idle for 30 minutes
APO Consumption : typical 10µA for Model 161 and 190µA for Model 162
Dimension : L227mm X W78mm X H40mm
Weight : 290 gm approx
Jaw opening & Conductor diameter : 50mm max
Accessories : Test lead pair, batteries installed, user's manual & soft carrying pouch
o
o
C to 40
C
o
C to 60
o
C, < 80% R.H. (with battery removed)
o
C
o
C @(0
10
o
C
o
C -18
o
C or 28
o
C

Advertisement