Other Settings

○Peak/Bottom value display
The maximum (minimum) pressure from when the power was supplied to this moment is detected and
updated. In peak/bottom display mode, this pressure is displayed.
For peak display, when the
starts flashing, and is held.
To release holding the display of the maximum pressure, press the
return to measurement mode.
For bottom display, when the
starts flashing and is held.
To release holding the display of the minimum pressure, press the
return to measurement mode.
If the
and
buttons are pressed simultaneously for 1 second or longer while the pressure is being held,
the maximum (minimum) values are reset.
○Zero clear
A displayed value can be adjusted to zero when the pressure to be measured is within ±7%F.S. (±3.5%F.S.
for compound pressure) of the pressure at the time of shipment from the factory.
(A tolerance range of ±1%F.S. may apply due to individual product differences).
Press the
and
buttons simultaneously for 1 second or longer, to clear the display to "0".
The display will return to measurement mode automatically.
For analogue output, the analogue output will be changed along with the display.
○Key lock
The key lock function is used to prevent errors occurring due to unintentional changes of the set values.
If a button operation is performed while the key lock setting is ON, "LoC" is displayed for approximately 1
second.
(When the
button is pressed, the set pressure will be displayed following "LoC").
<Operation - Without security code input- >
1, Press the
button for 5 seconds or longer in measurement mode.
The current setting "LoC" or "UnL" is displayed.
(Selecting or releasing the key lock is carried out in a similar way).
2, Press the
or
3, Press the
button to store the setting.
button is pressed for 1 second or longer, the maximum pressure and "Hi"
button is pressed for 1 second or longer, the minimum pressure and "Lo"
button to select key lock (or key unlock).
button for 1 second or longer again to
button for 1 second or longer again to
-44-