SMC Networks ZSE10(F) Operation Manual page 54

Digital pressure switch
Table of Contents

Advertisement

Fault No.
Fault
The pressure
display accuracy
12
does not satisfy
the specifications.
The display units
13
cannot be
changed.
The buttons
14
cannot be
operated.
The product is
15
loose.
The product is
16
noisy.
•Copy function
error
17
•Unable to copy
Possible cause
Check if any foreign matter has
Foreign matter
entered the pressure port.
Air or fluid
Check if air or fluid are leaking from
leakage
the piping.
Check if the product satisfies the
Insufficient
specified accuracy after 10 minutes
warm-up time
warm up time.
Product failure
Improper model
selection
Check if there is a "-M" at the end of
(Selection of
the part number printed on the
model "without
product
Units selection
function")
Product failure
Check if the key lock mode is turned
Key lock mode
on.
Product failure
Check that the panel mount adapter
Incorrect
and the product are correctly
installation
assembled.
Product failure
Air or fluid
Check if air or fluid are leaking from
leakage
the piping.
Product failure
(1) Check the wiring connection at the
Incorrect wiring
FUNC terminal.
(2) Check the power supply wiring.
Product failure
-53-
Investigation method
Countermeasure
Install a 5 µm filter to
prevent foreign matter from
entering the pressure port.
Also, clean the filter
regularly to prevent
drainage deposits.
Rework the piping.
If an excessive tightening
torque is applied, the
mounting bracket, screws
or the product may be
damaged.
After energizing, the display
and output can drift. For
detecting fine pressures,
warm up the product for 10 to
15 minutes.
Replace the product.
"M" in the part number
means that the unit cannot
be changed.
(kPa↔MPa can be
selected)
∗: The units selection function
is not available in Japan due
to a new measurement law.
∗: It is fixed to the SI unit "kPa",
"MPa".
Replace the product.
Turn off the key lock mode.
Replace the product.
Mount the product on the
panel correctly.
Replace the product.
Rework the piping.
If an excessive tightening
torque is applied, the
mounting bracket, screws
or the product may be
damaged.
Replace the product.
Correct the wiring.
Replace the product.