Advertisement
These safety instructions are intended to prevent hazardous situations and/or equipment damage.
These instructions indicate the level of potential hazard with the labels of "Caution", " Warning" or "Danger". They are all important notes for safety and must be followed in addition to International standards (ISO/IEC), Japan Industrial Standards (JIS) and other safety regulations.
| Operator error could result in injury or equipment damage. |
| Operator error could result in serious injury or loss of life. |
In extreme conditions, there is a possibility of serious injury or loss of life. |
Operator
|
|
|
|
|
|
|
|
|
NOTE
Names of individual parts
Indication light (Orange LED): Displays the switch operation condition.
LCD display: Displays the current status of pressure, setting mode and error code. Four display modes can be selected to display always in red or green only, or changing from green to red, red to green according to the output status.
button (UP): Selects the mode or increases the ON/OFF set value. Press this button to change to the peak display mode.
button (DOWN): Selects the mode or decreases the ON/OFF set value. Press this button to change to the bottom display mode.
button (SET): Press this button to change to either mode and to set a value.
*: The tube should be inserted to the end of the air-relieving port.
*: SMC TU0425 (polyurethane, O.D ø4, I.D ø2.5) is a suitable tubing.
Brown | DC(+) |
Gray | Analog output/Auto-shift input copy terminal |
Black | OUT1 |
White | OUT2 |
Blue | DC(-) |
Set ON point and OFF point of the Pressure switch.
Operation
When the pressure exceeds a set value, the Pressure switch will be turned on.
When the pressure falls below the set value by the amount of hysteresis or more, the Pressure switch will be turned off.
The default setting of the output set value is the central value between the atmospheric pressure and the upper limit of the rated pressure range. If the operation shown the right does not cause any problem, keep this operation setting.
<How to operate>
[Hysteresis mode]
[ Window comparator mode ]
The Pressure switch turns on within a set pressure range (from P1L to P1H) during window comparator mode. Set P1L (switch lower limit) and P1H (switch upper limit) using the setting procedure above.
Zero clear of indication Indication is reset to zero when ![]() ![]() For the first operation, perform zero clear without pressure supply. |
At the time of shipment, the following settings are provided.
If the setting is acceptable, keep it for use.
To change setting, refer to SMC website (URL http://www.smcworld.com) to get information in detail or contact us.
Unit specification | Pressure range | Default setting |
Nil or M | ISE40A | MPa |
ZSE40A(F) | kPa | |
P | ISE40A | PSI |
ZSE40A(F) |
Item | Explanation | Default setting |
Output mode | Selects hysteresis mode, window comparator mode or OFF mode. | Hysteresis mode |
Reversed output | Selects reversed output. | Normal output |
Pressure setting | Sets ON point or OFF point of the switch output | ISE40A: 0.500 MPa ZSE40A: -50.7 kPa ZSE40AF: 50.0 kPa |
Hysteresis | Chattering can be prevented by setting hysteresis. | 5% |
Display color | Selects the display color. | ON: Green OFF: Red |
Item | Default setting |
[F 3] Setting of response time | 2.5 ms |
[F 4] Setting of auto-preset | Manual |
[F 5] Setting of analog output/auto-shift input | Analog output |
[F 6] Setting of fine adjustment of display value | 0% |
[F11] Setting of display resolution | 1000-split |
[F80] Setting of power saving mode | OFF |
[F81] Setting of security code | OFF |
[F90] Setting of all functions | OFF |
[F97] Selection of copy function | OFF |
[F98] Check of output | Normal |
[F99] Reset to the default setting | OFF |
The measurement mode is the condition where the pressure is detected and indicated, and the switch function is operating.
This is the basic mode, and other modes should be selected for setting change and other function setting changes.
In measurement mode, press the button for 2 seconds or longer to display [F 0]. Select to display the function setting to be changed, [F
].
Press the button for 2 seconds or longer in function selection mode to return to measurement mode.
∗: Some functions are not available depending on part number. All functions are displayed with [F ] and followed with function description. If a function is not available for specified type, the function is displayed as [---].
To set each function the above in detail, refer to SMC website (URL http://www.smcworld.com) to get information in detail or contact us.
How to reset the product after power cut or forcible de-energizing
The setting of the product will be retained as it was before a power cut or de-energizing.
The output condition is also basically recovered to that before a power cut or de-energizing, but may change depending on the operating environment.
Therefore, check the safety of the whole facility before operating the product. If the facility is using accurate control, wait until the pressure switch has warmed up. (About 10 to 15 minutes)
This function is to display error location and content when a problem or an error occurs.
Error Name | Error Display | Error Type | Troubleshooting Method |
Over current Error | ![]() | A load current of switch output is 80 mA or more. | Turn the power off and remove the output factor for the over current. Then turn the power on. |
Residual Pressure Error | ![]() | During zero clear operation, pressure over ±7%F.S. (±3.5%F.S.for compound pressure) is applied. After 1 second, the mode will reset to the measurement mode. ±1%F.S. of the zero clear range varies with individual product differences. | Perform zero clear operation again after restoring the applied pressure to an atmospheric pressure condition. |
Pressurizing Error | ![]() | Pressure has exceeded the upper limit of the set pressure range. | Reset applied pressure to a level within the set pressure range. |
![]() | Pressure has exceeded the lower limit of the set pressure range. | ||
Auto-shift Error | ![]() | The measured pressure auto-shift input has exceeded the set pressure range. ∗: After 1 second, the mode will reset to measurement mode. | Auto-shift input signal is invalid. Check the connected equipment and correct the signal. |
System Error | ![]() | Displayed in the case of an internal data error. | Turn the power off and turn it on again. If resetting fails, an investigation by SMC CORPORATION will be required. |
If the error can not be reset after the above measures are taken, then please contact SMC.
Refer to the product catalogue or SMC website (URL http://www.smcworld.com) to get information about product specifications in detail.
Refer to the product catalogue or SMC website (URL http://www.smcworld.com) to get information about outline dimensions in detail.
Note: Specifications are subject to change without prior notice and any obligation on the part of the manufacturer.
© 2009 SMC Corporation All Rights Reserved
To get information in detail for operating this product, refer to SMC website (URL http://www.smcworld.com) or contact us.
Here you can download full pdf version of manual, it may contain additional safety instructions, warranty information, FCC rules, etc.
Download SMC Networks ZSE40A, ZSE40AF, ISE40A - Digital Pressure Switch Manual
Advertisement
Need help?
Do you have a question about the ZSE40A and is the answer not in the manual?
Questions and answers