Pre-Operation Checklist (Glow800) - Leica GLOW800 User Manual

Hide thumbs Also See for GLOW800:
Table of Contents

Advertisement

1
2
1 Microscope optics carrier
2 Test card
Eyepiece view:
Leica MFL800 Standard
10 746 513-01
0
1
2
3
!
Please read the Leica
MFL800 user manual!
Appearance of the test card in white light observation:
Holes in the colors squares will allow to check the adjustment of the
fluorescence and white light image.
4
5
6
7
8
9
GLOW800 / Ref. 10 747 241 / Version 00
Check the illumination and function & adjustement
9.3
Pre-operation checklist
(GLOW800)
Cleaning optical accessories
Check optical accessories for cleanliness.
X
X
Remove dust and dirt.
X
X
GLOW800 application
When using GLOW800 please ensure to have a Doppler
X
X
Ultrasound or similar device in place, in case of none or
insufficient blood flow visualization out of the ICG/GLOW800
procedure is given.
Balancing
Balance the microscope after refitting (see the user manual of
X
X
the Leica surgical microscope).
Operational check
Switch on the microscope.
X
X
Switch on the illumination.
X
X
Check the microscope illumination
X
X
Sterility
Apply the sterile drape.
X
X
For sterilizable components of the Leica surgical microscope
refer to the corresponding user manual.
21