Equipment required for installing the RKLC40-S scale

Required parts:
Appropriate length of RKLC40-S scale (see
A pair of standard 13 mm wide end clamps (A-9523-4015). Alternatively, a pair of 6 mm wide end clamps (A-9523-4111) are also available.
RGG-2 two-part epoxy adhesive (A-9531-0342)
Appropriate cleaning solvents (see
'Storage and handling'
RKLC40-S side mount scale applicator (A-6547-1912)
2 M2.5 screws
Green spacer (supplied with QUANTiC readhead)
Lint-free cloth
Optional parts:
Renishaw scale wipes (A-9523-4040)
Guillotine (A-9589-0071) or shears (A-9589-0133) for cutting RKLC40-S to length required
Magnet applicator tool (A-9653-0201)
Reference mark and limit magnets; see table below
Magnet type
Adhesive mounted magnets
Reference mark selector
1
Q limit
P limit
1
The reference mark selector magnet is only required for 'Customer selectable reference mark' readheads. For more information refer to QUANTiC
'RKLC40-S scale installation drawing'
on page 12)
on page 9)
Part number
Bolted magnets
(standard)
A-9653-0143
A-9653-0290
A-9653-0139
A-9653-0291
A-9653-0138
A-9653-0292
series encoder system data sheet (Renishaw part no. L-9517-9778).
www.renishaw.com