Set The Temperature Set Point - VWR International 1915A Installation And Operation Manual

Table of Contents

Advertisement

OPERATION
S
T
ET THE
EMPERATURE
Perform the steps below to adjust the set point to your process or application temperature.
1.
Set the Over Temperature Limit control to its maximum setting, if not already set to max.
2.
Navigate to the Temperature Set Point Adjustment mode
Briefly push and release either the Up or Down arrow
buttons to activate the temperature set point
adjustment mode.
Note: The display will automatically exit the adjustment mode after 5
seconds of inactivity, with the last shown set point value saved.
3.

Set the Temperature Set Point

Use the Up and Down arrow buttons to change the
temperature set point.
4.
Wait for 5 seconds after entering the Set Point
S
P
ET
OINT
This prevents the heating cutoff system from
interfering with this procedure.
The display will briefly flash the letters "SP",
then show the flashing, adjustable temperature
set point.
The display will stop flashing, and the set point
is now saved in the controller.
The chamber will now automatically heat or
passively cool to match your set point.
The display will revert to showing the current
chamber air temperature.
End of Procedure
Set Point Adjustment
Mode
Initial Set Point
New Set Point
Heating to New
Set Point