MAINTENANCE
Temperature Calibration Continued
7
Compare the reference device reading with the chamber
temperature display.
If the reference device and the chamber temperature
•
display readings are the same or the difference falls
within the range of your protocol, the incubator is
now calibrated for temperature.
See the next step if the readings fail to match or fall
•
outside of your protocol range.
8
If the difference still falls outside the acceptable range of
your protocol, repeat steps 3 – 7 up to two more times.
9
If the temperature readings of the incubator temperature display
and the reference device still fall outside your protocol after 3
calibration attempts, contact your distributor or technical
support for assistance.
-Or-
End of Procedure
Reference Device
Reference Device
37 |
P a g e