CONTROL PANEL OVERVIEW

Control Panel
Power Switch
Power is supplied and the switch illuminates when in the ( I ) ON position.
Set Temperature Display and Controls
Shows the current chamber temperature. The Up and Down arrow buttons are used to access
the Temperature Set Point (SP) or Calibration Offset (C O) display modes and input the
temperature set point or calibration adjustment value.
The Heating indicator light illuminates when the incubator powers the chamber heating elements.
Over Temperature Limit System (OTL)
This graduated dial sets the mechanical heating cut off point for the Over Temperature Limit
system. The system prevents unchecked heating of the chamber in the event of an electronics
failure or external heat spike. The red Over Temp Activated light illuminates when the OTL system
is rerouting power away from the heating elements. For more details, please see the
Over
Temperature Limit System
description in the Theory of Operations (page 25).