INSTALLATION
I
I
NSTALL THE
NCUBATOR
Install the unit in a workspace location that meets the criteria discussed in the previous entries of
the Installation chapter.
D
D
EIONIZED AND
Do not use deionized water to clean the unit, even if DI water is readily available in your
laboratory.
Use of deionized water may corrode metal surfaces and voids the manufacturing
warranty.
The manufacturer recommends the use of distilled water in the resistance range of 50K
Ohm/cm to 1M Ohm/cm, or a conductivity range of 20.0 uS/cm to 1.0 uS/cm, for cleaning
applications.
I
C
NSTALLATION
LEANING AND
The manufacturer recommends cleaning and disinfecting the shelving and chamber prior to
installation of the shelving in the chamber.
The unit was cleaned at the factory but may have been exposed to contaminants during
shipping.
Remove all wrappings and coverings from shelving prior to cleaning and installation. Do
not clean the shelving with deionized water.
Please see the
Cleaning and Disinfecting
chapter for information on how to clean and disinfect without damaging the unit.
W
ISTILLED
ATER
D
ISINFECTING
procedure on page 31 in the User Maintenance
19 |
P a g e