INSTALLATION
P
S
OWER
OURCE
When selecting a location for the unit, verify each of the following requirements is satisfied.
Power Source: The power source must match the voltage and amperage requirements listed on
the unit data plate. These units are intended for 110 – 120V, 50/60 Hz applications at the
following amperages:
Model
1915A
The wall power source must be protective earth grounded and single phase.
The unit may be damaged if the supplied voltage varies by more than 10% from the
data plate rating.
Use a separate circuit to prevent loss of the unit due to overloading or circuit failure.
The recommended wall circuit breaker is 15 amps.
The wall power source must conform to all national and local electrical codes.
Power Cord: The unit must be positioned so that all end-users can quickly unplug the cord in the
event of an emergency.
Each unit comes provided with a 125 volt, 15 Amp, 9 ft 5 in (2.89 m) NEMA 5-15P power
cord. Always use this cord or an identical replacement.
Fuses: These units each ship with a fuse installed in the power cord inlet.
The fuse must be installed and intact for the unit to operate.
Always find and fix the cause of a blown fuse prior to putting the unit back into operation.
Fuse type
o
250V T16A, 5x20mm
R
EQUIREMENTS
Amperage
12.0 Amps
Model
Amperage
14.5 Amps
1925A
Standard
NEMA 5-15R
wall socket
17 |
P a g e