INSTALLATION
P
S
OWER
OURCE
When selecting a location for the unit, verify each of the following requirements is satisfied.
Power Source: The power source must match the voltage and amperage requirements listed on
the unit data plate. These units are intended for 110 – 120V, 50/60 Hz applications at the
following amperages:
Model
1915A
The wall power source must be protective earth grounded and single phase.
•
The unit may be damaged if the supplied voltage varies by more than 10% from the
•
data plate rating.
Use a separate circuit to prevent loss of the unit due to overloading or circuit failure.
•
The recommended wall circuit breaker is 15 amps.
•
The wall power source must conform to all national and local electrical codes.
•
Power Cord: The unit must be positioned so that all end-users can quickly unplug the cord in the
event of an emergency.
Each unit comes provided with a 125 volt, 15 Amp, 9 ft 5 in (2.89 m) NEMA 5-15P power
•
cord. Always use this cord or an identical replacement.
Fuses: These units each ship with a fuse installed in the power cord inlet.
The fuse must be installed and intact for the unit to operate.
•
Always find and fix the cause of a blown fuse prior to putting the unit back into operation.
•
Fuse type
•
o
250V T16A, 5x20mm
R
EQUIREMENTS
Amperage
12.0 Amps
Model
Amperage
14.5 Amps
1925A
Standard
NEMA 5-15R
wall socket
17 |
P a g e