Laversab 6300-M4 User Manual page 56

Table of Contents

Advertisement

Note: If your Standard cannot generate 1.000 inHg abs., the lowest possible pressure below
4.000 inHg abs. will be acceptable. Enter the value of the generated pressure.
Step 8. On the keypad of the remote unit, press "PREV SCREEN". The display will return to
the Main Screen. If you press "CANCEL" the calibration adjustment will become void.
Step 9. Turn the 6300-M4 OFF, wait 30 seconds and then turn it back ON. The calibration
adjustment is now complete for the Pitot port.
Step 10. Verify the calibration by generating pressures in approximately 10% increments and
noting the 6300-M4 Pitot port (PtinHg) readings. Your calibration sheet will appear as shown
in Figure 6.1.
49

Advertisement

Table of Contents
loading

Table of Contents