6300-M4 display when this profile is executed.
Row 2 middle cell shows the number of the profile. This profile has a number 3. This tells
the 6300-M4 that this profile is to be loaded in as profile number 3. When profile 3 is
executed on the 6300-M4, this "Sample" profile will be executed. This number can be any
number between 1 and 20. When this profile is downloaded into the 6300-M4 it will
overwrite the previous profile that was resident in the 6300-M4 as number 3.
Row 2 right cell shows the height correction. This should normally be zero.
Rows 3 through 14 are the points of the profile. This profile has only 12 points. For each of
these points the columns contain the following information:
Column
A
B
C
D
E
F
G
In the sample profile, Row 3, which is the first point in the profile, tells the 6300-M4 to go in
Control mode to 100 knots on the Pitot side, and in Control mode to 0 feet at 6000 feet/min
on the Static side.
Row 4 (point 2) makes the 6300-M4 go to 200 knots and 0 feet at 6000 ft/min.
Row 5 (point 3) puts the Pitot side into Leak mode while still keeping Static in Control at 0
feet.
Row 6 (point 4) puts the Pitot side back into Control and takes it down to 100 knots. The
Static side is simultaneously moved up to 1000 feet at 6000 ft/min.
Row 7 (point 5) takes Pitot up to 200 knots and Static up to 5000 feet at 6000 ft/min.
Row 8 (point 6) takes Pitot up to 300 knots and Static up to 10,000 feet at 6000 ft/min.
Row 9 (point 7) takes Pitot down to 100 knots and Static up to 18,000 feet at 6000 ft/min.
Row 10 (point 8) keeps Pitot at 100 knots and puts Static into Leak mode while at 18,000
feet.
Row 11 (point 9) keeps Pitot at 100 knots and ramps Static in Control mode to 30,000 feet at
6000 ft/min.
Information
Pitot units
Pitot mode
Pitot target value
Static units
Static mode
Static target value
Rate target value
36