CyBio QuadStack
Basic information
Furthermore, the following applies in this context: The unfinished machine may only be
put into operation if it has been determined that the machine or installation into which
the unfinished machine is to be incorporated complies with the provisions of the direc-
tive 2006/42/EG and that a corresponding declaration of conformity in accordance with
annex II, part 1, section A has been issued.
The device meets the requirements of the directives 2014/30/EU and 2011/65/EU.
Any information regarding safety corresponds to the currently valid regulations of the
European Union. Other specific national laws and regulations must be observed.
Guard all hazard zones described in this manual for the safe integration of the device
into a machine or installation as far as possible from unintentional access. Preferably,
use a separating protective device to do this.
If the unfinished machine is incorporated into an external safety circuit via an E-stop ca-
ble, it is not permissible to have an E-stop jumper within the overall system.