Installation of the equipment

4.1

Electrical connection

The chambers are supplied ready for connection. They come with a fixed power connection cable of at
least 1800 mm / 70.87 in in length.
Model
BD all sizes
ED 23 / 53 / 115 / 240
FD all sizes
ED 400 / 720
BD 23-UL / 53-UL /
115-UL / 240-UL /
400-UL
ED 23-UL
FD 23-UL
BD 720-UL
ED 53-UL / 115-UL
FD 53-UL / 115-UL
ED 240-UL / 400-UL /
720-UL
FD 240-UL
• The domestic socket must also provide a protective conductor. Make sure that the connection of the
protective conductor of the domestic installations to the chamber's protective conductor meets the
latest technology. The protective conductors of the socket and plug must be compatible!
• Prior to connection and start-up, check the power supply voltage. Compare the values to the specified
data located on the chamber's type plate (chamber front behind the door bottom left-hand, or on the
left-hand side of the chamber, chap. 1.4).
• When connecting, please observe the regulations specified by the local electricity supply company
and as well as the VDE directives (for Germany). We recommend the use of a residual current circuit
breaker.
• Pollution degree (acc. to IEC 61010-1): 2
• Over-voltage category (acc. to IEC 61010-1): II
Danger of incorrect power supply voltage.
Damage to the equipment.
 Check the power supply voltage before connection and start-up.
 Compare the power supply voltage with the data indicated on the type plate.
See also electrical data (chap. 12.4 to 12.6).
To completely separate the chamber from the power supply, you must disconnect the power
plug. Install the chamber in a way that the power plug is easily accessible and can be easily
pulled in case of danger.
BD / ED / FD (E2) 03/2019
Nominal voltage ± 10% at the
Power plug
Grounded plug
CEE plug 5 poles
NEMA 5-15P
NEMA- 5-20P
NEMA L21-20P
indicated power frequency
230 V at 50 Hz
230 V at 60 Hz
400 V at 50 Hz
400 V at 60 Hz
115 V at 60 Hz
115 V at 60 Hz
208 V at 60 Hz
CAUTION
Current type
1N~
3N~
1N~
1N~
3N~
page 16/68