Model: 6300-M3
Full scale:
Ps: 38 inHg
Pt: 100 inHg
Last Calibrated date:
_______________________________________________________________________
TEST POINT
1. Vacuum
2. Fullscale
TEST POINT
1. Vacuum
2. Fullscale
_______________________________________________________________________
STATIC PRESSURE (Ps)
APPLIED
DISPLAYED
inHg
1.000
4.000
8.000
12.000
12.000
15.000
15.000
19.000
19.001
23.000
23.001
27.000
27.000
30.000
30.000
34.000
34.000
38.000
38.000
_______________________________________________________________________
Calibrated by:
10/10/01
CALIBRATION RESULTS
STATIC PRESSURE (Ps):
AS FOUND
AS LEFT
inHg
1.001
38.002
PITOT PRESSURE (Pt):
AS FOUND
AS LEFT
inHg
1.002
100.005
100.000
VERIFICATION RESULTS
inHg
1.000
4.000
8.000
Figure 6.1 Sample calibration sheet
56
Serial#: 74678
DEVIATION
inHg
1.000
38.000
DEVIATION
inHg
1.000
PITOT PRESSURE (Pt)
APPLIED
inHg
1.000
10.000
20.000
30.000
40.000
50.000
60.000
70.000
80.000
90.000
100.000
Date:
inHg
0.001
0.002
inHg
0.002
0.005
DISPLAYED
inHg
1.000
9.999
19.999
30.000
40.000
50.000
60.000
70.001
80.001
90.002
100.000