Heidolph Hei-PLATE Mix'n'Heat Expert Operating Instructions Manual page 123

Table of Contents

Advertisement

Available languages

Available languages

Appendix
IN_MODE_5\r\n
IN_MODE_5 Y\r\n
IN_MODE_6\r\n
IN_MODE_6 Y\r\n
IN_MODE_7\r\n
IN_MODE_7 Y\r\n
IN_MODE_8\r\n
IN_MODE_8 Y\r\n
Query status, type, version
Command
Response
STATUS\r\n
STATUS Y\r\n
S_CTRL\r\n
S_CTRL Y\r\n
S_OP\r\n
S_OP Y\r\n
SW_VERS V Y. Y.
SW_VERS\r\n
Y \r\n
M\r\n
M\r\n
T MR_
ULTIMATE\r\n
T\r\n
T MR_ EXPERT\r\n
QR_INFO\r\n
S#S#S\r\n
Query motor direction of rotation
• 0: CCW
• 1: CW
Query motor acceleration level
• 0: slow
• 1: medium
• 2: fast
Query ext. sensor monitoring
• 1: on
• 0: off
Query stir bar detector
• 1: on
• 0: off
Description
Status query
• Y = 0: Manual operation at the device
• Y = 1: Remote operation START_1/START_2
• Y = 2: Remote operation STOP_1/STOP_2
• Y < 0: Error code
• Y =-1: Remote blocked (device stopped manually)
Query current control source
• Y = 0: Manual, ext control allowed
• Y = 1: Profiles
• Y = 2: USB B
• Y = 3: RS-232
• Y = 4: LAN
• Y = 7: Manual, ext. control blocked
Query current control status
• Y = 0: All off;
• Y = 1: Heating on;
• Y = 2: Motor on;
• Y = 3: Heating and motor on
Y.Y.Y = HMI-Version (e.g: 1.2.345)
Identify device: Display of the tripped device begins
to flash
Query device type
Barcode string, format:
• Material number: max. 18 digits alphanum.
• Serial number: max. 12 digits alphanum.
• Production order: max. 8 digits numerical
• Separator: #
123