Electromagnetic Compatibility
The system is designed to be in compliance with IEC 60601-1-2, which contains electromagnetic
compatibility (EMC) requirements for medical electrical equipment. The limits for emissions and immunity
specified in this standard are designed to provide reasonable protection against harmful interference in a
typical medical installation.
The system complies with the applicable essential performance requirements specified in IEC 60601-1
and IEC 60601-2-18. Results of immunity testing show that the essential performance of the system is not
affected under the test conditions described in the following tables. For more information about the essential
performance of the system, see
Electromagnetic emissions
Table 14.
Guidance and Manufacturer's Declaration—Electromagnetic Emissions
The system is intended for use in the electromagnetic environment specified below. The customer or the
user of the system should ensure that it is used in such an environment.
EMISSIONS TEST
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Electromagnetic immunity
Table 15.
Guidance and Manufacturer's Declaration —Electromagnetic Immunity
The system is intended for use in the electromagnetic environment specified below. The customer or the
user of the system should ensure that it is used in such an environment.
IMMUNITY TESTS
Electrostatic discharge
(ESD)
IEC 61000-4-2
Operations & Maintenance Manual: Product Specifications
Essential Performance
COMPLIANCE
ELECTROMAGNETIC ENVIRONMENT – GUIDANCE
The system uses RF energy only for its internal function.
Group 1
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
Class A
The system is suitable for use in all establishments other than
Class A
domestic and those directly connected to the public low-voltage
power supply network that supplies buildings used for domestic
purposes.
In compliance
IEC 60601 TEST LEVEL
± 8 kV contact
± 16 kV air
on page 1.
COMPLIANCE
ELECTROMAGNETIC
LEVEL
ENVIRONMENT – GUIDANCE
Floors should be wood, concrete,
or ceramic tile. If floors are covered
In compliance
with synthetic material, the relative
humidity should be at least 30%.
53
0900‑5017 REV‑07