Verathon GlideScope Core Operation & Maintenance Manual page 55

Hide thumbs Also See for GlideScope Core:
Table of Contents

Advertisement

Table 9.
Core 10 FHD (0570‑0436)
Classification:
Line voltage:
DC power supply:
Ingress protection:
Battery type
Battery life
Charging time
Rated capacity
Nominal voltage
Nominal weight
Temperature:
Relative humidity:
Atmospheric pressure:
Display
Internal memory
Monitor (A)
Height (B)
Width (C)
Depth (D)
Weight
Operations & Maintenance Manual: Product Specifications
GENERAL SPECIFICATIONS
Electrical Class II, Applied Part BF
Range: 100–240 V AC, 50 and 60 Hz
(If the power cord has a third prong, the third conductor is not intended to be a
protective earth connection for the medical power supply.)
12 V DC, 2.5 A max
IP54
BATTERY SPECIFICATIONS
Lithium-ion
Under normal operating conditions, a fully charged, new battery lasts
approximately 135 minutes.
Charging time off-line will take no more than 4 hours from an empty battery to
a full charge.
3350 mAh
7.2 V
110 g (0.24 lbs)
OPERATING & STORAGE SPECIFICATIONS
Operating Conditions
10–35°C (50 to 95°F)
10–95%
700–1060 hPa
COMPONENT SPECIFICATIONS
TFT Color, 1920 x 1200 px
(MagnaView 1200x1200)
112 GB
25.7 cm (10.1 in)
206 mm (8.1 in)
269 mm (10.6 in)
48.5 mm (1.9 in)
1.25 kg (2.7 lb)
Shipping & Storage Conditions
-20–45°C (-4 to 113°F)
10–95%
440–1060 hPa
A
C
0900‑5017 REV‑07
B
D
49