Barcode Port; Lft (Sealed) - Quantronix Cubiscan 110-XT Operation And Technical Manual

Table of Contents

Advertisement

Communications

Barcode port

LFT (Sealed)

Cubiscan 110-XT
The scanner/barcode port can be enabled in the menus of the main
controller. The Expanded Communication Protocol is utilized when it is
enabled.
The barcode reader should be configured to send the barcode with a suffix
of <CR>.The Cubiscan uses the "End of Barcode" character as an indicator
that the barcode has been scanned in its entirety. Scanned barcodes are
transmitted with the next "Measure" command. If two barcodes are
scanned and then a measurement is taken, the most recent barcode will be
transmitted. If one barcode is scanned and then two measurements are
taken, the first measurement will contain the barcode while the second
measurement will contain no data for the barcode.
The bar code port is also a USB 2.0 type B connector, as shown below.
The following table shows the USB 2.0 type B connector pin assignments.
USB 2.0 type B Connector Pin Assignments
Pin
Name
1
VCC
2
D-
3
D+
4
GND
The Cubiscan is placed into LFT mode by moving two sided switches to the
"ON" position. This seals metrological parameters.
Cable
Description
Color
Red
+5 VDC
White
Data -
Green
Data +
Black
Ground
Barcode port
66