KaVo QUATTROcare PLUS 2124 A Instructions For Use Manual
KaVo QUATTROcare PLUS 2124 A Instructions For Use Manual

KaVo QUATTROcare PLUS 2124 A Instructions For Use Manual

Hide thumbs Also See for QUATTROcare PLUS 2124 A:

Advertisement

Quick Links

Instructions for use
QUATTROcare PLUS 2124 A – 1.008.3805

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the QUATTROcare PLUS 2124 A and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for KaVo QUATTROcare PLUS 2124 A

  • Page 1 Instructions for use QUATTROcare PLUS 2124 A – 1.008.3805...
  • Page 2 Distributed by: Manufacturer: KaVo Dental GmbH KaVo Dental GmbH Bismarckring 39 Bismarckring 39 88400 Biberach 88400 Biberach Germany Germany Phone +49 7351 56-0 www.kavo.com Fax +49 7351 56-1488...
  • Page 3: Table Of Contents

    Instructions for use QUATTROcare PLUS 2124 A – 1.008.3805 Table of contents Table of contents 1 User instructions....................... 1.1 User guide ........................5 1.1.1 Abbreviations....................1.1.2 General marks and symbols ................1.1.3 Target group ....................1.2 Service ........................6 1.2.1 Repair Service ....................
  • Page 4 Instructions for use QUATTROcare PLUS 2124 A – 1.008.3805 Table of contents 9 Terms and conditions of warranty ..................40 4 / 44...
  • Page 5: User Instructions

    Instructions for use QUATTROcare PLUS 2124 A – 1.008.3805 1 User instructions | 1.1 User guide 1 User instructions 1.1 User guide NOTE Read these instructions prior to first startup to avoid misuse and prevent damage. 1.1.1 Abbreviations International Electrotechnical Commission MPDG German Medical Device Law Implementation Act (Medizinprodukter- echt-Durchführungsgesetz, MPDG)
  • Page 6: Target Group

    If you have any technical questions or complaints, please contact the KaVo Technical service: +49 (0) 7351 56-1000 service.instrumente@kavo.com 1.2.1 Repair Service For repairs, please contact KaVo Repair Service. For scheduling or if you have any questions, please contact: KaVo Reparatur Service +49 (0) 7351 56-1900 E-mail: Service.Reparatur@kavo.com...
  • Page 7: Information On The Packaging: Storage And Transportation

    Terms and Conditions, Art. 28). Outside Germany NOTE KaVo shall not be held liable for damage arising from transportation. The shipment must be checked on arrival. If the packaging is visibly damaged on delivery, please proceed as follows: 1. The recipient of the package must record the loss or damage on the delivery receipt.
  • Page 8: Disposal

    Any waste produced must be recycled or disposed of in strict compliance with all applicable national regulations in a manner which is safe for both people and the environment. If you have any questions regarding the proper disposal of the KaVo product, please contact the KaVo branch office. 8 / 44...
  • Page 9: Disposal Of Electronic And Electrical Devices

    Instructions for use QUATTROcare PLUS 2124 A – 1.008.3805 1 User instructions | 1.5 Disposal of electronic and electrical devices 1.5 Disposal of electronic and electrical devices NOTE According to the general WEEE Directive (Waste Electrical and Electronic Equipment) and EU Directive 2012/19 concerning waste electrical and elec-...
  • Page 10: Safety

    4 Disconnect the product from the power supply when you leave your prac- tice. 2.3 Technical condition A damaged product or damaged or NOT KaVo original components could injure patients, users or third parties. A damaged power cable or missing protective conductor can lead to electrical shock.
  • Page 11: Accessories And Combination With Other Equipment

    NOTE If a repair is done with NON-KaVo original spare parts, this may constitute a product modification that leads to the loss of CE conformity. In the event of damage, the company performing the service or the operator is responsible.
  • Page 12 Instructions for use QUATTROcare PLUS 2124 A – 1.008.3805 2 Safety | 2.7 Material damage If compressed air is supplied without control, hoses on the inside of the product might burst or the compressed air controller might be damaged. 4 Before any extended period of non-use, disconnect the compressed air from the product.
  • Page 13: Device Description

    3 Device description 3.1 Intended use Indications for use: ▪ QUATTROcare PLUS 2124 A is used for the automated maintenance and re- moval of wear parts of dental instruments like handpieces, contra-angles and turbines in the dental therapy. ▪ A medical device according to relevant national statutory regulations.
  • Page 14: Quattrocare Plus 2124 A

    Instructions for use QUATTROcare PLUS 2124 A – 1.008.3805 3 Device description | 3.2 QUATTROcare PLUS 2124 A 3.2 QUATTROcare PLUS 2124 A ① 3x INTRAmatic service couplings, ⑤ Chuck service adaptor short ② 1x service coupling MULTIflex ⑥ QUATTROcare Plus Spraycan ③...
  • Page 15: Symbols On Device And Rating Plate

    Instructions for use QUATTROcare PLUS 2124 A – 1.008.3805 3 Device description | 3.3 Symbols on device and rating plate 3.3 Symbols on device and rating plate The rating plate is attached on the back of the unit. Accompanying documents...
  • Page 16: Technical Specifications

    Instructions for use QUATTROcare PLUS 2124 A – 1.008.3805 3 Device description | 3.4 Technical specifications 3.4 Technical specifications Device dimensions Width 350 mm (1.15 feet) Depth 230 mm (0.75 feet) Altitude 380 mm (1.25 feet) Weight 10 kg (22 pounds) Electrical supply Mains voltage 100 to 240 V AC...
  • Page 17: Scope Of Delivery

    Instructions for use QUATTROcare PLUS 2124 A – 1.008.3805 3 Device description | 3.5 Scope of delivery 3.5 Scope of delivery NOTE The package includes the articles listed below. If any of these articles is miss- ing, please contact the supplier immediately so that the missing article can be delivered.
  • Page 18 Instructions for use QUATTROcare PLUS 2124 A – 1.008.3805 3 Device description | 3.5 Scope of delivery CAUTION Accessories and combination with other equipment. Use of non-authorised accessories or non-authorised modifications of the device could lead to injury. 4 Only use accessories that have been approved for combination with the product by the manufacturer.
  • Page 19: Startup

    Instructions for use QUATTROcare PLUS 2124 A – 1.008.3805 4 Startup | 4.1 Selecting the location 4 Startup WARNING Explosion. Heating the device above 50°C (122°F) while the spray can is inserted can lead to explosions. 4 Protect the device from being heated above 50°C (122°F).
  • Page 20 Instructions for use QUATTROcare PLUS 2124 A – 1.008.3805 4 Startup | 4.1 Selecting the location WARNING Propellant gases in the QUATTROcare Plus Spray can. Explosive, flammable vapour/air mixtures may be produced during use. Risk of fire from overheating. 4 Follow instructions for use.
  • Page 21: Connecting The Quattrocare Plus 2124 A

    CAUTION Use of power cables other than original KaVo parts. Damage to or malfunction of the device. 4 Use original KaVo power cables only (for material numbers, see chapter on "Consumables and tools"). Rear of QUATTROcare PLUS 2124 A There are two fuses in the insert T1A250V ④.
  • Page 22: Inserting The Spray Can

    Instructions for use QUATTROcare PLUS 2124 A – 1.008.3805 4 Startup | 4.3 Changing the QUATTROcare Plus Spray can NOTE KaVo advises explicitly that original KaVo servicing products must be used ex- clusively. Also refer to: 7 Troubleshooting, Page 37 The QUATTROcare plus spray pack of 6 cans can be ordered under Mat. no.
  • Page 23 Instructions for use QUATTROcare PLUS 2124 A – 1.008.3805 4 Startup | 4.3 Changing the QUATTROcare Plus Spray can 4 Screw the spray can into the adaptor in clockwise direction and tighten until you notice a resistance. Do not jam it.
  • Page 24: Removing The Spray Can

    Instructions for use QUATTROcare PLUS 2124 A – 1.008.3805 4 Startup | 4.3 Changing the QUATTROcare Plus Spray can 4.3.2 Removing the spray can 4 Open the side door of the device. 4 Swing the spray can and the adaptor out of the device.
  • Page 25 Instructions for use QUATTROcare PLUS 2124 A – 1.008.3805 4 Startup | 4.3 Changing the QUATTROcare Plus Spray can NOTE The safety data sheet for the QUATTROcare Plus spray can be viewed at www.kavo.com. 25 / 44...
  • Page 26: Operation

    Instructions for use QUATTROcare PLUS 2124 A – 1.008.3805 5 Operation | 5 Operation WARNING Electrical shock. If a device or component is damaged, this may result in electric shock. 4 Only operate devices or components if they show no signs of damage on the outside.
  • Page 27: Control Panel

    Instructions for use QUATTROcare PLUS 2124 A – 1.008.3805 5 Operation | 5.1 Control panel 5.1 Control panel ① LED of spray can control ④ Start button for chucking system service ② LED of control light main switch ⑤ Start button for service procedure (on / off)
  • Page 28: Mounting Instruments

    Instructions for use QUATTROcare PLUS 2124 A – 1.008.3805 5 Operation | 5.3 Mounting instruments 5.3 Mounting instruments NOTE Remove the handpieces from the device as soon as the servicing is com- pleted. NOTE The LED display shows the following: Adaptor on the far left corresponds to LED display 1.
  • Page 29 Instructions for use QUATTROcare PLUS 2124 A – 1.008.3805 5 Operation | 5.3 Mounting instruments NOTE Carefully place the handpieces on the adaptor, taking care not to twist under pressure. 4 Mount the heads on the service coupling for INTRA heads (for special ac- cessories, see chapter on "Consumables and tools").
  • Page 30: Starting The Servicing Procedure

    Instructions for use QUATTROcare PLUS 2124 A – 1.008.3805 5 Operation | 5.4 Starting the servicing procedure 5.4 Starting the servicing procedure NOTE When the LED of the spray can control ① lights up, the spray can is empty. When the main switch is on, the LED ② lights up.
  • Page 31: Repositioning The Service Couplings

    Instructions for use QUATTROcare PLUS 2124 A – 1.008.3805 5 Operation | 5.6 Repositioning the service couplings 4 Pull the high speed air handpiece in the axial direction of the service coup- ling, and take it off. 4 Press on the latch with your thumb and take the straight and contra-angle handpiece off by pulling it carefully in the direction of the axis of the service coupling at the same time.
  • Page 32: Performing The Chucking System Service

    Instructions for use QUATTROcare PLUS 2124 A – 1.008.3805 5 Operation | 5.7 Performing the chucking system service 4 Unscrew the union nut clockwise in the direction of the arrow, and take the service coupling off. 4 Attach the service coupling as selected and fasten it turning the union nut counterclockwise against the direction of the arrow.
  • Page 33: Setting The Blow-Out Time

    ð The spray is then briefly injected through the chucking system, completing the service. End the chucking system service mode: 4 Place the handpieces in the QUATTROcare PLUS 2124 A, close the front flap and start the service procedure. ð Alternatively, after three minutes with no service procedure running, the device automatically switches back to normal service mode.
  • Page 34 Instructions for use QUATTROcare PLUS 2124 A – 1.008.3805 5 Operation | 5.8 Setting the blow-out time 4 Hold down the start key ②. 4 Turn the main switch ③ on. ð LEDs 1 to 4 of the LED display ① light up in succession.
  • Page 35: Maintenance

    Instructions for use QUATTROcare PLUS 2124 A – 1.008.3805 6 Maintenance | 6.1 Cleaning 6 Maintenance CAUTION Improper maintenance and repair. Risk of injury. 4 The device should be serviced, cleaned, and stored in a dry location, and should be disconnected from the mains, according to instructions, if it is not to be used for extended periods.
  • Page 36: Safety Check

    KaVo Customer Service Centre Warthausen or a test centre approved by KaVo. The QUATTROcare PLUS 2124 A must be subjected to a service check including safety check every 5 years. The service check may only be done by a profes- sional trained by KaVo or in a shop trained by KaVo.
  • Page 37: Troubleshooting

    7 Troubleshooting NOTE If a repair is done with NON-KaVo original spare parts, this may constitute a product modification that leads to the loss of CE conformity. In the event of damage, the company performing the service or the operator is responsible.
  • Page 38: Consumables

    Instructions for use QUATTROcare PLUS 2124 A – 1.008.3805 8 Consumables 8 Consumables Presentation Material summary Mat. no. Service coupling 1.011.7380 INTRAmatic, short Service coupling MUL- 1.009.6142 TIflex Service coupling for IN- 0.411.7941 TRA heads and actuating sleeve Service coupling 1.005.1707...
  • Page 39 Instructions for use QUATTROcare PLUS 2124 A – 1.008.3805 8 Consumables Presentation Material summary Mat. no. Service coupling for NSK 1.000.8786 high speed air handpieces Service coupling for MOR- 1.000.6063 ITA high speed air hand- pieces BIEN-AIR service coup- 1.001.7860 ling W&H service coupling...
  • Page 40: Terms And Conditions Of Warranty

    12 months from the date of purchase as confirmed by the sales- person. In case of justified complaints, KaVo will honour its warranty with a free re- placement or repair. The warranty does not cover defects and their consequences that arose or may...

This manual is also suitable for:

1.008.3805

Table of Contents