Agilent Technologies U1271A Service Manual page 42

Handheld digital multimeter
Hide thumbs Also See for U1271A:
Table of Contents

Advertisement

1
Calibration Procedures
Using the Front Panel for Adjustments
26
Table 1-3 Adjustment input values (continued)
Test function
Step
30 mA
300 mA
ACmA/ACA
3 A
10 A
OPEN
10 nF
100 nF
Capacitance
1000 nF
10 μF
100 μF
1000 μF
10 mF
Reference value
Valid reference input
0.9 to 1.1 × Reference value
03.000 mA (70 Hz)
0.9 to 1.1 × Reference value
30.000 mA (70 Hz)
0.9 to 1.1 × Reference value
030.00 mA (70 Hz)
0.9 to 1.1 × Reference value
300.00 mA (70 Hz)
0.9 to 1.1 × Reference value
0.3000 A (70 Hz)
0.9 to 1.1 × Reference value
3.0000 A (70 Hz)
0.9 to 1.1 × Reference value
3.0000 A (70 Hz)
0.9 to 1.1 × Reference value
10.000 A (70 Hz)
OPEN
OPEN terminals
0.9 to 1.1 × Reference value
04.000 nF
0.9 to 1.1 × Reference value
10.000 nF
0.9 to 1.1 × Reference value
010.00 nF
0.9 to 1.1 × Reference value
100.00 nF
0.9 to 1.1 × Reference value
0100.0 nF
0.9 to 1.1 × Reference value
1000.0 nF
10.000 μF
0.9 to 1.1 × Reference value
100.00 μF
0.9 to 1.1 × Reference value
1000.0 μF
0.9 to 1.1 × Reference value
0.9 to 1.1 × Reference value
10.000 mF
U1271A/U1272A Service Guide

Advertisement

Table of Contents
loading