Agilent Technologies U1271A Service Manual page 41

Handheld digital multimeter
Hide thumbs Also See for U1271A:
Table of Contents

Advertisement

U1271A/U1272A Service Guide
Table 1-3 Adjustment input values (continued)
Test function
Step
3 V
30 V
ACV
300 V
1000 V
OPEN
300 μA
DCµA
3000 μA
300 μA
ACμA
3000 μA
OPEN
30 mA
DCmA/DCA
300 mA
3 A
10 A
Calibration Procedures
Using the Front Panel for Adjustments
Reference value
Valid reference input
0.9 to 1.1 × Reference value
0.3000 V (70 Hz)
0.9 to 1.1 × Reference value
3.0000 V (70 Hz)
0.9 to 1.1 × Reference value
3.0000 V(3 kHz)
0.9 to 1.1 × Reference value
3.000 V (70 Hz)
0.9 to 1.1 × Reference value
30.000 V (70 Hz)
0.9 to 1.1 × Reference value
30.000 V(3 kHz)
0.9 to 1.1 × Reference value
30.00 V (70 Hz)
0.9 to 1.1 × Reference value
300.00 V (70 Hz)
0.9 to 1.1 × Reference value
300.00 V (3 kHz)
0.9 to 1.1 × Reference value
30.0 V (70 Hz)
0.9 to 1.1 × Reference value
300.0 V (70 Hz)
0.9 to 1.1 × Reference value
300.0 V (3 kHz)
OPEN
OPEN terminals
300.00 μA
0.9 to 1.1× Reference value
3000.0 μA
0.9 to 1.1 × Reference value
030.00 μA (70 Hz)
0.9 to 1.1 × Reference value
300.00 μA (70 Hz)
0.9 to 1.1 × Reference value
300.0 μA (70 Hz)
0.9 to 1.1 × Reference value
3000.0 μA (70 Hz)
0.9 to 1.1 × Reference value
OPEN
OPEN terminals
0.9 to 1.1 × Reference value
30.000 mA
0.9 to 1.1 × Reference value
300.00 mA
0.9 to 1.1 × Reference value
3.0000 A
0.9 to 1.1 × Reference value
10.000 A
1
25

Advertisement

Table of Contents
loading

This manual is also suitable for:

U1272a

Table of Contents