Introduction and Description

Characteristics

Several rotors with a wide range of tubes can be used.
The set speed is reached within seconds. The
maintenance-free induction motor ensures quiet and low
vibration operation even at high speeds, and guarantees a
very long lifetime.
The user-friendly control panel makes it easy to preset the
speed, RCF value, running time, temperature, and running
profile (acceleration and braking curves). You can choose
between the display of speed and RCF or the entry mode.
These settings can be changed even while the centrifuge is
running.
With the help of the PULSE key
a sample for just a few seconds, if necessary.
The Fisherbrand GT 2 / 2R Centrifuge is equipped with
various safety features:

Technical Data

The technical data of the Fisherbrand GT 2 / 2R Centrifuge is listed in the following table:
Feature
Environmental conditions
Permissible ambient temperature
Overvoltage category
Pollution degree
Heat dissipation
IP
Running time
Max Speed n
max
Min Speed n
min
Centrifuge GT 2 / 2R
, you can also centrifuge
PULSE
Values
Use in interior spaces
Altitudes of up to 2000 m above sea level
Max. relative humidity 80 % up to 31 °C; decreasing linearly up to 50 %
relative humidity at 40 °C.
+2 °C to +35 °C
II
2
Refrigerated
230 V
4 778 BTU/h
20
Unlimited
15 200 rpm (depending on the rotor)
300 rpm
The housing and rotor chamber consist of steel plate and
the interior is made of armour steel, while the front panel
is made of high-impact resistant plastic.
The lid is equipped with a view port and a lock.
The lid of the centrifuge can only be opened while the
centrifuge is switched ON and the rotor has come to a
complete stop. The centrifuge cannot be started until the
lid is closed properly.
The integrated rotor detection system ensure that no
inadmissible speed settings can be preselected.
Electronic imbalance recognition
Lid emergency release: For emergencies only, e.g. during
power failures (Refer to the section Mechanical
Emergency Door Release)
+2 °C to +35 °C
II
2
Ventilated
230 V
3 447 BTU/h
20
Unlimited
15 200 rpm (depending on the rotor)
300 rpm