Respironics SmartMonitor 2 Service Manual page 59

Table of Contents

Advertisement

Verification of Low Breath Rate Alarm
Test Point
Simulator Setting
1
Resp. Breaths/Min
2
Resp. Breaths/Min
Verification of the Low and High Heart Alarms
Test Point
Simulator Setting
1
ECG Beats/Min
2
ECG Beats/Min
Press the RESET button to reset the red Low Heart Light
3
ECG Beats/Min
4
ECG Beats/Min
Press the RESET button to reset the red High Heart Light
Due to large signal changes that result when Simulator settings are initially changed, the off-scale signal recognition circuitry of
the SmartMonitor 2 may activate and cuase brief pauses in detection. Once normal signal detection by the monitor resumes,
allow an additional 10 seconds for the circuitry to stabilize before proceeding with verifying the operation of the
SmartMonitor 2.
1007897
Detection
Low Breath Alarm
10
YES
Within 20 seconds
(+/-5), the red
APNEA light should
flash and the audible
alarm should beep
once per second.
50
YES
Alarm Low Heart
Red Light
On.
50
YES
YES
100
NO
YES
240
YES
NO
100
NO
NO
SmartMonitor 2 Service Manual
Pass
NO
High
Pass
Heart Red
Light On.
NO
NO
YES
YES
Chapter 8: Verification
Simulator Setting for Test Set Up
ECG Beats/Min
ECG AMPL./MV
Variation OHMS
Resp. Breaths/Min
Base Impedance OHMS
Simulator Setting for Test Set Up
ECG Beats/Min
ECG AMPL./MV
Variation OHMS
Resp. Breaths/Min
Base Impedance OHMS
100
0.5
1
50
200
100
0.5
1
50
200
8-12