Modem Auto Dial
Plug the telephone wire into the modem connector on the
back of the SmartMonitor .
Plug the other end of the phone wire into the wall phone
jack.
Refer to the Synergy-E™ software for instruction on receiving a
call from the SmartMonitor 2.
Modem will continue calling every 30 minutes, for 4 hours until it
connects with the computer. If no connection is made, the modem
will try again the next day at the pre-selected time. For example, if
the modem is programmed to call at 7am but fails to make a
connection after 8 attempts it will stop trying. The next day at 7am
it will try again. This will continue until a connection is made.
Modem Auto Answer in communications Mode
SmartMonitor 2 Service Manual
Chapter 4: Installation and Setup
Connect a phone cord from the wall jack to the modem connector on
the back of the SmartMonitor 2.
To allow the SmartMonitor 2 to work with the modem, place the monitor
in Communications Mode:
Press and hold the blue RESET button.
Press and release the POWER button.
Wait until the monitor alarms. Now release the RESET
button.
Briefly press and release the RESET button again.
The SmartMonitor 2 beeps every 10 seconds whenever it
is in the Communications Mode. This is a reminder that the
SmartMonitor 2 is powered on for working with the
computer, memory card, or modem, but not for monitoring the
infant.
Do nto connect the infant to the SmartMonitor 2 when in the
Communications Mode, the spnea and heart rate alarms are not
operational in this mode.
The dealer should now call to retrieve the information. The
phone may ring, but do not answer. The modem answers
the call and connects the SmartMonitor 2 to the computer
through the phone line.