Troubleshooting Table (Continued)
Symptom
All lights will flash together and the alarm
will beep in unison with the flashing lights.
Pressing RESET will not silence alarm.
(continued)
Alarm sounds weak.
Loose lead
Light remains on.
1007897
Possible Cause
Incorrect Power-Off Sequence.
No power, battery drained.
Internal part failure.
Internal part failure.
Low battery
Connections between electrodes,
lead wires, and patient cable are
not properly made.
Defective lead wires.
Defective electrodes.
Internal part failure.
SmartMonitor 2 Service Manual
Chapter 6: Troubleshooting and Diagnostics
Corrective Action
Press the POWER button and ensure that
the power light is illuminated.
Press and hold the RESET button. Press
and release POWER. Wait two seconds,
then release the RESET button.
Connect power cord/battery charger. Use Power-Off
to silence alarm.
Press and hold the blue RESET button.
Press and release the gray POWER button.
Wait 2 seconds, then release the RESET Button.
Prior to use, allow battery to charge 6 hours.
Contact Respironics.
Contact Respironics.
Charge battery
Verify that
a. babys skin underneath electrodes is clean.
b. electrodes are clean and lead wires are fully
inserted into the electrodes and patient cable.
Replace lead wires and perform Functional Self-
Test.
Replace electrodes.
Contact Respironics
6-4