To Set a specific parameter or series of parameters:
Use the UP arrow button to move forward to the parameter to be changed.
Select the parameter to be changed by pressing the ENTER key - this will cause the current value to flash
The value can be increased by pressing the UP arrow button and decreased by pressing the DOWN arrow button..
To select the displayed value, press the ENTER button. The value displayed will continue to flash until you press the ENTER button.
Press the UP arrow button to proceed to the next parameter to be selected.
Press the UP arrow to display CLEAR MEMORY?, then clear the memory.
For the purposes of this check out procedure, set the SmartMonitor 2 parameters to the following values.
Parameter
Patient Name
Patient ID number
STD Alarm Parameters
STD Record Parms
STD System Parms
Apnea Alarm
Low Breath Rate Alarm
Bradycardia Alarm
Bradycardia Alarm Delay
Tachycardia Alarm
Tachy Alarm Delay
Record Mode
Apnea for Record
Brady for Record
Pre/Post Time
Impedance Record?
Record Respiration Rate
1007897
Value
Test
1234
Are Selected
Not Selected
Not Selected
20 Seconds
OFF
80 BPM
5 seconds
230 BPM
0 seconds
EVENT
16 Seconds
90 BPM
45/15
YES
YES
SmartMonitor 2 Service Manual
Parameter
Heart Rate Record
ECG Record
Auxiliary 1 Record
Auxiliary 2 Record
Auxiliary 3 Record
Auxiliary 4 Record
External Physiological Trigger
External Equipment Trigger
Date
Time
Rate Display
Phone Number for Computer
Time to Call the Computer
Call Computer when Memory Is Full
Move Data to Card
Memory Status
Clear Memory
Chapter 8: Verification
Value
YES
YES
OFF
OFF
OFF
OFF
OFF
OFF
Current Date
Current Time
ON
(___-____)
(__/__/__ __:__)
OFF
NO
0 Percent Full
NO
8-4