KROHNE OPTIWAVE 7300 C Quick Start Manual page 18

Non-contact radar (fmcw) level meter for liquids
Hide thumbs Also See for OPTIWAVE 7300 C:
Table of Contents

Advertisement

INSTALLATION
2
• Make sure the tank connection is level.
• Make sure that you use the applicable gasket for the connection dimensions and the process.
• Align the gasket correctly.
• Lower the antenna carefully into the tank.
• Turn the nut on the device process connection to attach the device to the tank.
• Tighten the connection.
i
Refer to local rules and regulations for the correct torque to apply to the connection.
SMS
Equipment needed:
• Device
• Gasket (not supplied)
• SMS nut
Figure 2-20: SMS connection
1 Tank socket
2 Nut for SMS connection
• Make sure the tank connection is level.
• Make sure that you use the applicable gasket for the connection dimensions and the process.
• Align the gasket correctly.
• Lower the antenna carefully into the tank.
• Turn the nut on the device process connection to attach the device to the tank.
• Tighten the connection.
i
Refer to local rules and regulations for the correct torque to apply to the connection.
18
www.krohne.com
OPTIWAVE 7300 C
04/2010 - 4000050003 - QS OPTIWAVE 7300 R06 en