6 Technical specifications
Parameter
Type of device
Sample volume
Measuring range
Measuring principle
Test duration
Measurement temper-
ature range
Humidity
Storage temperature
Specification
Lactate Scout Vet - hand-held device for measuring blood lactate
0.2 μL
0.5 mmol/L - 25.0 mmol/L
Enzymatic amperometric determination of lactate using
lactate oxidase
10 seconds
+10 °C - +45°C
10 - 85% relative humidity (max. 2 minutes for Test Strips)
-18 °C - +8 °C (Test Strips in container) and/or
-20 °C - +50 °C (device)
34
Need help?
Do you have a question about the SensLab Lactate Scout Vet and is the answer not in the manual?