Errors/warnings
Description and potential causes
No acoustic signal
No acoustic signal in any mode
Volume was muted in the "Configuration
and Display" menu.
Device turns itself off
Device turns
itself off
The display shows a bright EKF logo on
a black background.
This happens automatically after two
minutes of inactivity.
Ambient temperature is too low or
batteries are empty.
Display malfunctioning.
Defective electronics or
mechanical damage.
Defective
Values seem too high/too low
test results
Perspiration, intended puncture site
not washed correctly or perspiration
occurred due to a delay between
washing and collecting the blood.
Intended puncture site was still wet
after washing.
Solution
Open the "Volume" menu and change
the settings.
Restart the measuring device.
Make sure that the ambient temperature
is within the operating range of the
measuring device.
Change the batteries.
Please contact the
EKF Technical Support team.
Repeat the test with a new Test
Strip. Check the code configuration.
Precisely follow the instructions for
performing the test.
Carefully wash perspiration away from
the intended puncture site and dry your
hands well.
Use the check solution to perform a
functionality test. Please contact the
EKF Technical Support team if there is
an error message.