3.4 Functionality test

The functionality test is used to check the
correct functioning of the Lactate Scout Vet
measuring system. If there are any doubts
regarding the accuracy of the test result or the
correct functioning of the measuring device,
you must perform a functionality test. Please
use the Lactate Scout Check Solution for
the functionality test. The label on the Check
Solution bottle states the range of the Lactate
Scout Check Solution.
The check solution is available in the following
concentrations:
8.9 - 1 1.1 mmol/L (display: 10 mmol/L)
4.5 – 5.6 mmol/L (display: 5 mmol/L)
The Check Solution can be used for a period
of three months after initial opening for the
first time. Once opened, ensure the check
solution is stored tightly closed between
15 – 25°C.
For functionality test, Check Solution, Test
Strip and measuring device must be at the
same temperature.
ATTENTION
Never bring the opening of the Lactate Scout
Check Solution bottle directly into contact
with the Test Strip. Failure to do so will
contaminate the contents of the bottle
and render it unusable.
WARNING
Do not swallow the Lactate Scout Check
Solution.
Avoid contact with mucous membranes.
Risk of small parts being ingested:
keep the Lactate Scout Check Solution out of
the reach of children.