ATTENTION
The Test Strip and the Lactate Scout Vet must
be at the same temperature when performing
a test. If you take the Test Strip container
out of the refrigerator, you must wait for at
least 20 minutes until the Test Strip container
has reached the ambient temperature before
you opening the container. If you take the
Test Strip container out of the freezer,
you must wait for at least 2 hours until
the Test Strip container reaches ambient
temperature. Only remove Test Strips to be
used immediately for testing purposes. Reseal
the container immediately after removing
Lactate Scout Test Strips
After removing Lactate Scout Test Strips
from the container measurement of the Test
Strips must be accomplished within 2 minutes.
Please do not store Lactate Scout Tests Strips
for more than 2 minutes outside the storage
container or at relative humidity >85%.
2.2 Blood collection and test
Wash the sample site with clean water. Dry the
sample site. In order to take a capillary blood
sample, puncture the site using a sterile lancet.
Wipe away the first droplet of blood. Apply
gentle pressure to the puncture site. The
second droplet must be large enough to fill
the measurement chamber of the Test Strip
in one go.
WARNING
Avoid pressing on the puncture site too
heavily as sweat and/or tissue fluid can merge
with the droplet of blood and interfere with
the test result.
The droplet of blood must not run. Position
the tip of the Test Strip on the sample site
and the Test Strip will draw blood into the
measuring chamber