2.3 Product overview
The 9525sc degas-only DCCP (degas cation conductivity panel) system removes CO
degassed cation conductivity (DCC). The system measures a sample stream of high purity water.
The system has one sensor and one controller. Refer to
Figure 1 9525sc degas-only DCCP system
1 Heater
2 Sample cooler
3 Polymetron 8315 conductivity sensor
4 SC200 Controller (or Polymetron
9500 controller)
5 Electrical enclosure
6 Power switch
1
The power indicator light starts and stops with the heater.
Figure
1.
7 Power indicator light
8 Temperature controller
9 Sample flow meter
10 Sample flow valve
11 Flow cell
and measures
2
1