Specifications - Keysight 478A Operating And Service Manual

Option h83 thermistor mount manual modifications
Hide thumbs Also See for 478A:
Table of Contents

Advertisement

478A Option H83

Specifications

Modify the 478A Operating and Service Manual (00478-90021),
TABLE 1 "SPECIFICATIONS" to the following specifications.
Table 1
Electrical Specifications
Parameter
Frequency Range
Impedance
Max Return
Loss/SWR
Operating Resistance
Power Range
Max Peak Power
Max Average Power
Max Energy per Pulse
RF connector
Bridge Connector
The 478A Option H83 comes with calibration data for 1, 3, 10, 30 ,50, 100, 300, and 1000
MHz from a Standards Calibration Lab, which is traceable to NIST.
All other specifications in the 478A Operating and Service Manual, Table 1 apply except
where 431 use is referenced in the standard manual.
NOTE
The 476A Option H83 can not be used with the 431 Power Meters.
Thermistor Mount Manual Modifications
Specification
Condition
1 MHz to 1 GHz
50 Ω
17.7 dB / ≤ 1.3
1 MHz to < 50 MHz
32.3 dB / ≤ 1.05
50 MHz
17.7 dB / ≤ 1.3
> 50 MHz to 1 GHz
200 Ω ± 1%
µ
Cannot use 431 Power Meter
1
W to 10 mW
200 W
30 mW
µs
pulse rep rate ≥ 1 kHz
10 w /
pulse rep rate < kHz
µs
5 w /
Type N (m)
Mates with cable
supplied with 432
type Power
Meters
General Information
Specifications
3

Advertisement

Table of Contents
loading