System Calibration - GasTech D-Guard2 Instruction Manual

Hide thumbs Also See for D-Guard2:
Table of Contents

Advertisement

D-Guard² Instruction Manual
21.2 System Calibration
You can adjust the D-Guard² 4-20mA current loop output.
Access the Advanced Settings menu. See
section
21.1.
Step 1. Use
to go into the advanced menu.
Step 2. Use
to go into system calibration.
Step 3. Use
or
to go into the menu.
Step 4. Use
to set then
to save.
NOTE: The Output Count provides visual feedback as the output is adjusted.
Step 5. Use
to go back or
to continue.
42