Auto Span Calibration - GasTech D-Guard2 Instruction Manual

Hide thumbs Also See for D-Guard2:
Table of Contents

Advertisement

D-Guard² Instruction Manual
15.6 Auto Span Calibration
Access the calibration menu. See
Step 1. Use
to go into the auto span mode.
Step 3. Use
to save or
CAUTION: You must monitor the alarm states before you exit the calibration menu, as alarms
may remain active until the calibration gas has been removed from the sensor.
Step 5. Use
to exit or
NOTE: The D-Guard² Auto Calibration mode determines when the rate of change of the
applied gas has decreased to less than
WARNING! Incorrect calibration can result in injury or death to personnel.
20
section
15.1.
to cancel.
to cancel.
± 0.1% of full scale per second.
Step 2. Apply the correct gas. Use
Step 4. Use
to exit or
Step 6. Calibration completed.
to start.
to continue.
Item Number 201018

Advertisement

Table of Contents
loading

Table of Contents