SECTION 4 – TROUBLESHOOTING GUIDE
4-2
Symptom
Sample fails prematurely.
Serial communication problems.
Reason
Initial sample load is excessive.
Sample was not placed in the
center of the platen.
Incorrect cable.
Controller or display is not
configured for serial
communication.
Action
Check the adjustment of the top
platen limit switch. Adjust the
top platen control switch to
close before the platen touches
the top of the press. Verify the
time delay relay is set to approx.
20 seconds.
Relocate the sample and repeat
the test.
Verify the serial communication
connections.
Verify address and connection
parameters (Baud Rate and
Parity) match between
Controller and 5270