CHANDLER
ENGINEERING
TITLE: Procedure, 5270 Configuration for 4207D
Instrument Model: Model 4207D Compressive Strength Tester
Revision
Date
B
10/01/09
C
02/23/23
The Model 4207D Compressive Strength Tester may be used with the Model 5270 Data Acquisition and Control
System (DACS). The following procedure is used to configure 5270 for use with this instrument. The following
features are available:
•
Acquisition of peak value from the controller
•
Acquisition of the programmed value from the controller
•
Download of API Fast and Slow load rates to the controller
•
Automatic reset of the load display peak value when a new 5270 test is started
Using Tools > Configure > I/O Connections, configure an I/O connection using the Modbus protocol.
Select the COM port that is connected to the 4207D instrument. COM4 is illustrated; however, the actual port
assignment may vary as required by the hardware configuration.
Revised By
AMH
WJW
Description
ECN T2532
ECN T9560; EPC Update
Part Number: 89-0175
Page 1 of 2
Checked By
TC
JS